Background

The ability to detect and quantify methane concentrations has importance in both environmental and process safety contexts. For example, the ability to detect natural gas leaks on offshore platforms is a vital part of explosion prevention schemes. A methane cloud larger than 5m in diameter is large enough to cause damage if ignited and, consequently, detectors may need to be placed on a grid with spacing no greater than 5m, leading to deployment of large numbers of sensors. In order to operate in explosion prevention regimes, the sensors must detect methane gas below the lower explosion limit so that operators can take action before an explosive mix of air and methane is created by any potential leak. For example, a common action point is at 20% of the lower explosion limit (LEL) and therefore a sensor needs to be able to reliably detect methane concentrations of this magnitude. As the LEL of methane is 4.9% by volume, a sensor needs to be able to detect and quantify methane concentrations of 1% by volume or 10,000ppm, with a precision of around 500ppm.

Typically on an oil platform, or petrochemical plant, large numbers of sensors will be placed at different locations, which is expensive due to the cost of the sensors themselves and the need for installation and maintenance within explosion hazard zones. As well as methane, sensors are needed that respond to leaks of other potentially hazardous gases such as hydrogen sulphide and oxygen (in areas where asphyxiants are used). Pellistor sensors are typically used to detect potentially explosive gases, but these require regular calibration because of known effects of drift and poisoning. Electrochemical sensors are well established for potentially toxic gases and oxygen, but have a limited lifetime.

For these reasons, interest has been growing in detection technology based on tunable diode laser spectroscopy (TDLS), with its high sensitivity, calibration stability and specificity to particular gases with no cross-response (for example, to humidity). However, the cost of this technology is prohibitive when each detection point requires its own tunable laser. One solution to the problem has been to share the light from a single laser across multiple sensor points connected via optical fibre. However, this solution requires a complicated network of optical fibre, with an output and return fibre needed for every detection location, thus 2n individual fibres connect to n separate detection points as well as an individual photo detector and associated digital acquisition channel for each sensor. Our work has been to multiplex several sensor locations along a single optical fibre, using a single laser diode and photo detector, thereby significantly simplifying both the installation and interrogation unit and providing a lower cost per sensor location, such that the technology becomes competitive with traditional point sensors.

Technology overview

¹û½´ÊÓƵ¹ÙÍø has developed a method to enable simultaneous measurements of gas concentrations at multiple points, using an optical fibre, with a single fibre-coupled laser and photodetector, by combining tunable diode laser spectroscopy (TDLS) and range resolved interferometry (RRI). Closely spaced gas detection cells can be deployed along the length of the fibre and can be positioned to follow an individual gas pipe or to provide multi-point gas sensing adjacent to an individual process or other hazardous location. Detection locations can be as close as 1m apart along the fibre, or can be spaced further apart, limited by the coherence length of the laser (lasers with long coherence lengths of up to 2km are commercially available). In principle, we believe our solution can be implemented as connectorised sections, enabling different configurations with a modular product. Improved sensitivity and coverage of gas detection are enabled.

 

Above: RRI interrogation unit.

The principle of operation of TDLS permits the gas detection through measurement of the optical absorption of gas at multiple sampling points, while the wavelength of a diode laser is tuned through an absorption line. Range resolved interferometry (RRI) on the other hand enables separation of signals from each sensor by employing a special signal processing technique, previously developed for position sensing and vibrometry.

The emission wavelength of the diode laser is controlled by the injection current and modulated with a combination of slow ramp (2Hz) and fast sinusoidal (49kHz) wave forms before coupling into a fibre. The slow ramp sweeps the central wavelength of the laser through a range containing a characteristic gas absorption, e.g. for methane at 1653nm. The RRI signal processing then demodulates the spatial location of the output signal to provide a TDLS-like signal for each gas cell.

We have demonstrated that signals from three gas cells can be independently distinguished and measured with minimal cross-talk. Depending on the optical topology and response time, methane can be measured with a noise-equivalent concentration between 20 and 200ppm. In other work, RRI has been combined with other measurands to multiplex 20 sensors along a single optical fibre.

Compared to standard TDLS, the principle innovation is the use of RRI signal processing. For RRI only a commercial field-programmable gate array (FPGA) processor, a standard laser driver and single photodetector are required. We use standard telecommunications fibre with low cost connectors and couplers. Optical alignment of the gas cells is eased by using a retroreflector design.

Benefits

The new type of multi-point optical gas detection system features a single light source and detector multiplexed across many cells and offers significant advantages over existing sensors and networked sensor systems, including:

  • Potential to multiplex a large number of cells for widespread coverage;
  • Close spacing of detection points for explosion prevention (with >1m fibre separation between cells);
  • Ability to measure methane at low noise equivalent concentrations of less than 200ppm for a 0.5s measurement period for early detection (and a potential noise equivalent concentration [1σ] of <20ppm with 1500s averaging times);
  • Minimal cross-talk between cell measurements for sensitivity and localisation;
  • Higher gas specificity and signal-to-noise ratio compared with non-dispersive infrared measurements to prevent false alarms;
  • Simultaneous measurement and reporting of all cell concentrations, assisting real-time gas mapping;
  • Low power lasers (<35 mw) compared to multi-point photo-acoustic gas detectors, for safe operation in explosion-prone environments (optical power complies with EN 60079-28:2015);
  • Single system-in-a-box controller and single connecting fibre cable for convenient installation and accessibility for maintenance;
  • Potential for much lower multi-point sensor costs per measurement point;
  • Improved lifetime and lower maintenance.

The system can be translated (via a change of laser) to enable detection of a range of gases as well as methane, including hydrogen sulphide, carbon monoxide and carbon dioxide, ethane, propane and ethylene and others (with detectable absorbances within the transmission window of standard telecommunications fibres).

Applications

The multi-point distributed optical system for gas detection can be used in established gas detection applications in industrial safety (e.g. explosion protection on petrochemical works, toxic gas monitoring), landfill monitoring and process monitoring.

Opportunity

¹û½´ÊÓƵ¹ÙÍø has a track record of delivering research and commercial projects related to novel optical instrumentation and sensors, to solve challenging measurement problems in engineering and biomedical applications, including microfluidic flows, point-of-care diagnostics, strain, pressure and gas detection and multi-wavelength gas spectroscopy.

We are seeking expressions of interest from industrial partners wishing to utilise the know-how on design, simulation, control and operation of novel multi-point gas sensing systems. Licensees and potential users of the technology are invited to get in touch.

Patents

UK (GB1814542.5) and PCT applications (WO2020/049287) are in progress with a priority date of 6 September 2018.

IP status

  • Real-time, laboratory bench-based experimental demonstration completed with three independent gas cells;
  • RRI established for other measurands in challenging environments (e.g. installed on a rotating helicopter rotor blade);
  • Different measurement topologies evaluated.

Seeking

  • Licensing,
  • Development collaboration, for example, via application to Innovate-UK for co-funding.